Operating Information
AM700 Audio Measurement Set Service Manual
2–25
NOTE. Calibrate does not perform an analog acquisition system calibration
when Digital Interface Tester is the running application.
The Conversion Standards menu shown in Figure 2–20 permits you to define
certain values to be used in deriving measurement results.
Output dBm. This setting defines the value of the external load resistor. The
divider formed by load resistance and the internal series resistance selected for
use (10, 150, or 600 W) is taken into account for the calculation (see
Figure 2–21). Thus, generator signal level can be specified in dBm, and the
correct power will be delivered to the load.
NOTE. This is calculated dBm, so if the external load is not as stated in the
Output dBm box, the output value stated in dBm will not be correct.
Figure 2–20: Configure system setup conversion standards menu
System Conversion
Standards Menu