Maintainence
6–10
AM700 Audio Measurement Set Service Manual
5. Use the large control knob to select C, then press the Select front-panel
button to enter your selection. This brings up the CLEAR SYSTEM
NVRAM Utility.
CLEAR SYSTEM NVRAM UTILITY
=========================
### WARNING -- Zeroing system NVRAM will destroy: ###
--- All user files: setups, functions,
GPIB addr, COMM port settings, etc
--- Instrument saved state settings
--- Instrument console info
--- Diagnostics Error log
--- Custom Diagnostic sequences
After NVRAM has been cleared, the instrument will be reset to
the Diagnostics Startup Menu, and Diagnostics NVRAM will be
re-initialized. The instrument's NVRAM will be reset to its
factory default configuration when the instrument itself is
run.
### Press - MENU BUTTON - to zero NVRAM, any other to abort ###
6. After the NVRAM is zeroed, you may use the menu choices and Select
button to back out of the menu and select an application to start from the
Instrument – RUN MODES – Menu (a choice in the AM700 STARTUP
MENU) or you may turn the AM700 off and back on again to restart the
applications.