Adjustment Procedures
AM700 Audio Measurement Set Service Manual
5–31
6. After the calibration has finished, press Clear Menu. Then, select Digital as
the application to load.
1. With the Digital Application running, press the Configure front panel button
and again touch the System Setup soft key.
2. Select Calibrate and touch the Calibrate Now soft key.
3. After the calibration has finished, press Clear Menu.
This completes the adjustment procedure for the AM700.
Functional Check
You can perform the functional checks in the Verification Procedure to verify the
operation of the instrument.
Digital Calibration