Tektronix AM700 Stereo System User Manual


 
Diagnostics
AM700 Audio Measurement Set Service Manual
6–19
High Level Diagnostic Interface
--- INSTRUMENT RUN MODES ---
Key Selection
__________________________________________________________
A ... -AUTORESET- ........ (R) run all diagnostics
Q ... -QUICK START- ...... (R) skip all diagnostics
---------------------------------------------------------
= ... Reprint @ ... Abort Menu
---------------------------------------------------------
Input your selection:
L ... -LOW LEVEL MENU- ... (R) run Low Level Diags Infc
NOTE: Modes marked (R) will RESET the instrument.
N ... -NORMAL- ........... (R) run powerup diagnostics
Figure 6–8: Instrument run modes in V1.0/V1.01 firmware
The AM700 then resets and displays the AM700 STARTUP MENU choices
shown in Figure 6–9.
Low Level Diagnostic Interface
--- AM700 STARTUP MENU ---
Key Selection
__________________________________
H ... Diagnostics -HELP- Menu
R ... Instrument -RUN MODES- Menu
D ... Low Level -DIAGNOSTICS- Menu
__________________________________
= ... Reprint @ ... Abort Menu
__________________________________
Input your selection:
U ... -UTILITIES- Menu
F ... -FIRMWARE- Date Codes
Figure 6–9: AM700 startup menu
The second method of accessing the Low Level diagnostics menu starts with the
power off. This method is referred to as the “backdoor boot modes.” Press and
hold the Menu button while turning on the power. After the AM700 beeps twice
to acknowledge the startup mode, release the Menu button. The AM700 then
initializes directly into the Low Level diagnostics mode as shown in Figure 6–9.