Performance Verification
4–8
AM700 Audio Measurement Set Service Manual
Amplitude and frequency
knob assignment icons
Figure 4–4: Generator control display
5. Check that the analog generator is set to High Resolution, Ch A is set to
Sine, Ch B is set to Follow A, and both generator outputs are enabled as
shown in Figure 4–4. Also check that the control knobs for setting the
generator frequency and amplitude are assigned to the analog generator.
6. Touch the Waveform Control icon on the left side of the screen to display the
Waveform Control screen as shown in Figure 4–5.
7. Set the Analog A generator (Ch A) to output a frequency of 1 kHz by
touching the Frequency edit box to highlight it. Enter a 1 on the numerical
key pad and then kHz in the units keys for 1 kHz.
8. Touch the Amplitude edit box to highlight it. Enter a 0 on the numerical key
pad followed by dBu in the units keys for an amplitude of 0.0 dBu.