Tektronix AM700 Stereo System User Manual


 
Diagnostics
AM700 Audio Measurement Set Service Manual
6–31
Exception errors occurring during step 1 (Address space probe) will:
Display an error message sequence on the Diagnostic LED display,decoded
as follows:
An “E” will precede the exception number (0–255).
A “P” will precede the Probe step number (see Table 6–9).
Numbers are shown as a sequence of decimal digits, separated by
decimal points (to allow detecting repetition of digits), most significant
digit to least significant.
For example, a bus error (exception 2) occurring during probe step 15
would display the following sequence:
E", 2", P", 1", .", 5".
If the error is considered critical:
Jump to the startup code (warm reboot).
Otherwise:
Continue testing
Exception errors occurring during steps 2 through 16 will:
Display an error message sequence on the Diagnostic LED display, 2 times
through.
The error message is decoded as follows:
An “E” precedes the exception number (0–255).
A dash/minus sign precedes the boot step number.
Numbers are shown as a sequence of decimal digits, separated by
decimal points (to allow detecting repetition of digits), most significant
digit to least significant.
For example, a divide by zero error (exception 5) occurring during boot
step 15 would display the following sequence twice through.
E", 5", Ć", 1", .", 5"
Jump to the startup code (warm reboot).
Exception errors occurring after step 16 will:
Display an error message on the instrument screen (or to the serial port).
Wait for user acknowledgement.