6. Set the AM700 (DUT) under test as follows:
a. Press the GENERATOR/Control button, touch the DSP soft key, and
then touch the Waveform Control soft key. Make the following
selections:
Select Digital SF 1
Sine
Frequency = 1 kHz
Amplitude = 0 dBu
Select Digital SF 2
Sine
Frequency = 5 kHz
Amplitude = 0 dBu
b. Press the Clear Menu button.
c. In the GENERATOR control panel, press the Generator On button.
d. Press the FFT button, and then press the Configure button.
e. Touch the System Setup soft key, and then touch the Conversion
Standards soft key. Make the following selection:
Digital Audio Sampling Rate = Interface Clock
f. Touch the Accept Changes soft key.
7. Set the the test AM700 as follows:
a. Press GENERATOR/Control button
Touch DSP soft key
Touch Waveform Control soft key
b. Select Digital SF 1
Sine
Frequency = 10 kHz
Amplitude =0 dBu
c. Select Digital SF 2
Sine
Frequency = 15 kHz
Amplitude = 0 dBu
d. Press Clear Menu button
e. Press GENERATOR On button