Performance Verification
4–6
AM700 Audio Measurement Set Service Manual
Analog Generator Checks
Initial checks list actual button presses and screen touches (soft keys) in detail to
assist you in becoming familiar with the user interface operation of the AM700.
As you progress through this procedure, some assumptions about your famil-
iarity are made to reduce the instructions. As an example you may see the
instruction to “make sure the generator is in High Bandwidth mode” where it is
assumed that you know how to do that without explicit instructions.
1. If the AM700 is not on, turn it on and wait until it initializes with the FFT
application running. If already on, skip to step 2.
2. Press the Configure button and touch the Input Selection soft key to display
the Input Configuration control panel as shown in Figure 4–2.
Figure 4–2: Input configuration control panel
NOTE. Ch 1 and Ch 2 are measurement channels of the AM700. Any of the input
channels of the AM700 may be connected to either or both measurement
channels using the Input Selection controls.
Analog Generator High
Resolution and
High Bandwidth
Mode Operation