Performance Verification
4–22
AM700 Audio Measurement Set Service Manual
Figure 4–14: Audio Analyzer graphical display with real time (RT) view visible
12. Touch the number area in the Range edit field of Ch A, press the Units
button in the keypad (see Figure 4–15), and press the V/Hz button to select
V as the units for the input range.
Figure 4–15: Keypad entry buttons
13. Use the key pad to set the input range to 3.88 V.
14. Repeat step 12 and step 13 for Ch B.