Tektronix AM700 Stereo System User Manual


 
Firmware Upgrade Procedure
6–92
AM700 Audio Measurement Set Service Manual
10. Follow the instructions that appear on the AM700 screen to complete the
reprogramming procedure.
11. Press Enter or Sweep/Run to continue, or press Clear Menu to stop the
upgrade procedure. You will need Disk #1 and Disk #2 ready to insert when
asked to do so.
Each disk is read and verified before the firmware is upgraded. If an error
occurs while reading the upgrade data disks, the upgrade procedure is
terminated without changing the loaded firmware.
12. When both data disk have loaded the upgrade data a CRC is calculated, and
you will be given the choice of programming the Flash EEPROM or
aborting without affecting the stored data.
13. Press Sweep/Run to continue programming the Flash EEPROM. The old
code is erased and the new code is uploaded. Loading time is approximately
seven minutes.
14. Press Sweep/Run or Clear Menu to reboot the AM700 and run the upgraded
firmware.
15. After the AM700 has rebooted and is running, press the Configure button.
16. Touch the State Saving soft key to display the menu used to either save state
or to restore factory defaults.
17. Select Restore Factory defaults and confirm that you want to restore the
factory defaults. This loads the new factory defaults that are modified in the
new firmware.