Performance Verification
AM700 Audio Measurement Set Service Manual
4–21
Figure 4–13: Input Configuration control panel
4. Press the Clear Menu button.
5. Press the Control button in the Generator section of the front panel.
6. Set the A analog generator to High Res mode, with a sine-wave signal output
as shown in Figure 4–4 on page 4–8.
7. Set the B analog generator to Follow A mode. Touch the Waveform Control
soft key and set Analog A Frequency to 1.00 kHz and Amplitude to 3.50 V.
Press the Clear Menu button.
8. Turn on the RT display window and turn off the other display windows as
shown in Figure 4–14.
9. Set the Level Units to Volts to match the generator units for amplitude
settings.
10. Press the Menu button, and touch the Input soft key to display the Input
Configuration menu.
11. In the Input Configuration menu, set the input ranges to manual for Ch A
and Ch B by deselecting the Auto range in the Input Configuration control
panel (shown in Figure 4–13). (The check mark is removed from the ✓ Auto
choice, and the manual Range edit fields are not grayed out.)