To set up the ICP supply test
Performance Test and Operation Verification
This test verifies that the Agilent 35670A meets its input specification for ICP
signal conditioning. In this test, a digital multimeter directly measures the open
circuit voltage of each channel. The digital multimeter measures the current souce
of each channel by measuring the voltage across a 50
feedthrough termination.
The digital multimeter is set to the 100 V range to measure open circuit voltage and
set to the 1 V range to measure the current source.
1 2ch
2 2ch
Agilent 35670A Verifying Specifications
To set up the ICP supply test
3-41