Time Domain
Specifications apply in histogram/time mode, unfiltered time display
DC amplitude accuracy
±5.0%fs
Rise time of 1 V to 0 V test pulse
<11.4 ms
Settling time of 1 V to 0 V test pulse
<16 ms to 1%
Pulse aberrations (peak overshoot)
of 1 V to 0 V test pulse
Peak aberration relative to the mode-to-mode
difference (most common values)
<3 %
Sampling period
1 channel mode
2 channel mode
4 channel mode (option AY6 only)
3.815 ms (1/262144 Hz) to2sin2× steps
7.629 ms (1/131072 Hz) to4sin2× steps
15.26 ms (1/65536 Hz) to8sin2× steps
Trigger
Trigger modes Internal trigger
External trigger
Source trigger
GPIB trigger
Maximum trigger delay
Post trigger
Pre trigger
No two channels can be further than ±7168 samples
from each other.
8191 seconds
8191 sample periods
External trigger maximum input
±42 Vpk
External trigger range
Low range
High range
2Vto+2V
10 V to +10 V
External trigger resolution
Low range
High range
15.7 mV
78 mV
Agilent 35670A Specifications
Time Domain
1-9