Agilent Technologies 35670-90066 Stereo System User Manual


 
To run the program without a printer
Use this procedure if you do not have an GPIB printer connected to yout system.
Write in the information needed on the title page of the selected test record.
The test records are located near the back of this chapter and may be copied without
written permission of Agilent Technologies.
Press the following keys and when the program prompts you, type in the model
number and GPIB address:
[
EQUIP CONFIG ]
[
AC CALIBRATO ]
[
SYNTH. 1 ]
[
SYNTH. 2 ] (If needed)
[
LOW-D OSCILLATO ] (If needed)
[
MULTIMETER ]
[
RETURN ]
The GPIB addresses equals 100
(interface select code) + (primary address). The
interface select code for the test equipment is 7 (for example, if the primary address is
8, the GPIB address is 708).
Press the following keys:
[ TEST CONFIG ]
[
PROCEDURE ]
[
OP_VERIFY ]or[PERFORMAN ]
[
STOP AFTER ]
[
EACH MEASUREMENT ]
[
RETURN ]
Press the following keys to start the test:
[
START TESTING ]
[
START BEGINNING ]
Verifying Specifications Agilent 35670A
To run the program without a printer
3-10