Agilent Technologies 35670-90066 Stereo System User Manual


 
To set up the source amplitude accuracy test
Performance Test and Operation Verification
This test verifies that the Agilent 35670A meets its source output specification for
sine amplitude accuracy at 1 kHz. In this test, a digital multimeter measures the
amplitude accuracy of the source. Source amplitude accuracy is checked at
0.1 Vpk with the digital multimeter set to the 100 mVrms range and at 3.0 and
5.0 Vpk with the digital multimeter set to the 10 Vrms range. For the standard two
channel analyzer, the digital multimeter is connected to the rear panel source
connector instead of the front panel source connector. This is the only test that
verifies the rear panel source port on the standard two channel analyzer.
1 2ch
1 4ch
Agilent 35670A Verifying Specifications
To set up the source amplitude accuracy test
3-45