Agilent Technologies 35670-90066 Stereo System User Manual


 
Press the following keys:
[
System Utility ]
[
MORE ]
[
SERVICE TESTS ]
[
ADJUSTMTS ]
[ ADC ADJUSTMNT
]
[
OFFSET ]
Wait for the analyzer to set up the adjustment. The analyzer is ready when the
adjustment message appears on the screen.
If the oscilloscope display looks like the following figure, go to step 17.
The following describes the signals shown on the oscilloscope display:
‘’
A’’ A straight, horizontal trace in the upper half of the display.
‘’B’’ A “noisy” flat trace at the center of the sine wave trace.
‘’C’’ A clean sine wave in the lower half of the display.
R431 and R405 Correctly Adjusted
If trace ‘’B’’ is not flat, adjust A5 R431.
If trace ‘’C’’ is not centered over trace ‘’B,’’ adjust A5 R405.
Set the power switch to off (
O ).
Disconnect the jumper from A5 TP8 and A5 TP300. Reconnect the cable to
A5 P4.
Agilent 35670A Adjusting the Analyzer
To adjust the ADC gain, offset and reference
5-9
A
B
C