To set up the amplitude accuracy test
Performance Test and Operation Verification
This test verifies that the Agilent 35670A meets its single channel amplitude
specification for FFT full scale accuracy at 1 kHz. In this test, an ac calibrator
outputs a 1 kHz signal with an exact amplitude to all channels. This test checks
amplitude accuracy at 27, 19, 9, 1, −11, −27, −35, −43, and −51 dBVrms.
1 2ch
1 4ch
Agilent 35670A Verifying Specifications
To set up the amplitude accuracy test
3-17