Agilent Technologies 35670-90066 Stereo System User Manual


 
Set the oscilloscope as follows:
Channel 1
Volts/Div
Offset
Coupling
20 mV/div
0V
1M
ac
Channel 2
Volts/Div
Offset
Coupling
500 mV/div
0V
1M
ac
Time Base Time/Div
Sweep
1.0 ms/div
Triggered
Trigger Source
Level
Slope
Mode
Channel 2
500 mV
Positive
Edge
Display Mode
Averaging
No. of Avg.
Screen
Repetitive
On
8
Single
Remove the cable from A5 P4. Connect a jumper from A5 TP8 to A5 TP300.
Press [
Preset ] while setting the power switch to on ( I ).
Press the following keys:
[
System Utility ]
[
MORE ]
[
SERVICE TESTS ]
[
ADJUSTMTS ]
[
ADC ADJUSTMNT ]
[
SECOND PASS GAIN ]
Wait for the analyzer to set up the adjustment. The analyzer is ready when the
adjustment message appears on the screen.
Adjust A5 R407 for a flat trace on the oscilloscope.
Set the power switch to off (
O ).
Disconnect the capacitive load and connect the 1:1 oscilloscope probe directly
to the oscilloscope input. Attach the probe to A5 TP402 and the probe ground
clip to the instrument chassis.
Change the set up for the oscilloscope as follows:
Channel 1
Volts/Div
Coupling
115 mV/div
1M
dc
Channel 2
Coupling
1M
dc
Time Base
Time/Div 500 ms/div
Display Averaging
Display time
off
2.00 s
Set the power switch to on ( I ).
Adjusting the Analyzer Agilent 35670A
To adjust the ADC gain, offset and reference
5-8