If the signal at A5 TP 204 is toggling, the A6 Digital assembly is probably
faulty.
•
If the signal at A5 TP 204 is not toggling, the A5 Analog assembly is
probably faulty.
Agilent 35670A Troubleshooting the Analyzer
To troubleshoot trigger failures