Now follow the directions on the display and record every measurement result
in the selected test record.
Warning During the test, the program prompts you to change the test equipment
connections. Always turn the ac calibrator output to OFF or STANDBY before
changing test equipment connections. The ac calibrator can produce output
voltages that could result in injury to personnel.
The directions on the display briefly tell you how to connect test equipment. For
detailed illustrations of equipment setup, see the setup illustrations starting on page
3-13.
If you want to pause the program and return the Agilent 35670A Dynamic Signal
Analyzer to front panel control, press [
BASIC ]. To continue the program, press
[
BASIC ][DISPLAY SETUP ][LOWER ][RETURN ][CONTINUE ]. If you changed
any instrument setup states, press [
RESTART TEST ] instead of [ CONTINUE ]to
ensure accurate measurement results.