A12 BNC
The A12 BNC assembly is only used in two channel analyzers. The A12 BNC
assembly routes the signals connected to the Channel 1 BNC connector and Channel 2
BNC connector to the A1 Input assembly. The A12 BNC assembly also routes the
source signal from the A5 Analog assembly to the Source BNC connector. The signal
connected to Channel 1 should be between 0.19531 Hz to 102.4 kHz in single channel
mode and between 0.097656 Hz to 51.2 kHz in two channel mode. The signal
connected to Channel 2 should be between 0.097656 to 51.2 kHz. For both channels,
the signal’s amplitude range must be between +27 dBVrms and –51 dBVrms for full
scale measurements. The source output signal can be random noise, burst random
noise, periodic chirp, burst chirp, pink noise, or fixed sine. The amplitude range of the
source output is
±5 Vpk.
Voltages and Signals Agilent 35670A
A12 BNC