† The sync output to signal output phase error was determined to be less than 0.25 degrees.
‡ If measured value is within 3% of specification, verify synthesizer level accuracy. Note: Without 50 termination,observed levels
are twice the setting into high impedance.
Performance Test
Using Recommended Test
Equipment
Using Other Test Equipment
Measurement
Uncertainty
Ratio
Measurement
Uncertainty
Ratio
Source Amplitude Accuracy
0.1 Vpk
3.0 Vpk
5.0 Vpk
± 9.83 mVpk
± 492.9 mVpk
± 633.0 mVpk
>10:1
>10:1
>10:1
Source Output Resistance
±0.15
>10:1
Source DC Offset
0 Vdc, 0 Vac-pk
±10 Vdc, 0 Vac-pk
0 Vdc, 5 Vac-pk
±5 Vdc, 5 Vac-pk
± 238 nV
±84mV
± 2.123 mV
± 43 mV
>10:1
>10:1
>10:1
>10:1
Source Flatness
± 0.2 dB
5.24:1
Source Distortion
fundamental <30 kHz
fundamental 30 kHz
±0.5 dB
±0.5 dB
6.3:1
>10:1
Verifying Specifications Agilent 35670A
Measurement Uncertainty
3-58