Agilent Technologies 35670-90066 Stereo System User Manual


 
Specifications
This chapter contains the specifications for the Agilent 35670A Dynamic
Signal Analyzer and the critical specifications for the equipment required to
test the Agilent 35670A.
Instrument specifications apply after 15 minutes warm-up and within 2 hours of
the last self-calibration. When the internal cooling fan has been turned OFF,
specifications apply within 5 minutes of the last self-calibration. All
specifications are with 400 line frequency resolution unless stated otherwise.
Four channel instruments are unspecified in the one channel mode where alias
protection filters are not connected.
Abbreviations
dBVrms = dB relative to 1 Volt rms.
dBfs = dB relative to full scale amplitude range. Full scale is approximately 2 dB below ADC
overload.
FS or fs Full scale; synonymous with input range.
Real Time or Online = Refer to the collecting and displaying of information with no dropouts
or missing information.
Rload = Load resistance connected to the analyzer’s source.
Typical = Typical, non-warranted, performance specification included to provide general
product information.
Vpk = Peak of the ac voltage.