Agilent Technologies 35670-90066 Stereo System User Manual


 
Verifying Specifications
This chapter tells you how to use the Agilent 35670A Semiautomated Performance
Test Disk
. The performance test disk contains a program that semiautomates the
operation verification tests and performance tests.
After you review this chapter, follow the directions in ‘’To load the program’’
then continue with one of the following:
‘’To run the program in semiautomated mode’’
‘’To run the program without a printer’’
‘’To run the program in manual mode’’
Caution Before applying line power to the analyzer or testing its electrical performance, see
chapter 2, ‘’Preparing the Analyzer for Use.’’
Overview
The
Semiautomated Performance Test Disk contains a program (ITM_35670A)
and two procedure files (OP_VERIFY and PERFORMAN). ITM_35670A is
the test manager program. OP_VERIFY is the operation verification procedure
file and PERFORMAN is performance test procedure file. The procedure files
contain an ordered list of tests, and each test contains one or more
measurements. Since ITM_35670A reads the procedure files, the disk must
remain in the disk drive during testing.
If you do not have a keyboard connected to the analyzer, use the numeric key
pad and the alpha keys when the program prompts you to type in information.
See the analyzer’s help text for a description of the alpha keys.
If a test fails, contact your local Agilent Technologies sales and service office
or have a qualified service technician see chapter 4, ‘’Troubleshooting the
Analyzer,’’ in the
Agilent 35670A Service Guide.
3-2