Agilent Technologies 35670-90066 Stereo System User Manual


 
Half Range A In both the two channel and four channel analyzer, a low on this line turns on
the channel 1 half range LED. This line goes low when the A1 or A2 Input assembly detects
that the amplitude of the channel 1 input signal reached half the set range. This line is HRNGA
inverted by the A11 Keyboard Controller assembly.
HRNGBn Half Range B — In a two channel analyzer, a low on this line turns on the channel 2 half range
LED. In a four channel analyzer, a low on this line turns on the channel 3 half range LED.
This line goes low when the A1 Input assembly detects that the amplitude of the channel 2
input signal reached half the set range or when the A2 Input assembly detects that the
amplitude of the channel 3 input signal reached half the set range. This line is HRNGB
inverted by the A11 Keyboard Controller assembly.
HRNGCn Half Range C — In a four channel analyzer, a low on this line turns on the channel 2 half range
LED. This line goes low when the A2 Input assembly detects that the amplitude of the channel
2 input signal reached half the set range. This line is HRNGC inverted by the A11 Keyboard
Controller assembly. This line is only used in four channel analyzers.
HRNGDn Half Range D — In a four channel analyzer, a low on this line turns on the channel 4 half range
LED. This line goes low when the A2 Input assembly detects that the amplitude of the channel
4 input signal reached half the set range. This line is HRNGD inverted by the A11 Keyboard
Controller assembly. This line is only used in four channel analyzers.
OVLDAn Overload A — In a four channel analyzer, a low on this line turns on the channel 1 overload
LED. This line goes low when the A2 Input assembly detects that the amplitude of the channel
1 input signal exceeded the set range. This line is only used in four channel analyzers.
OVLDBn Overload B — In four channel analyzers, a low on this line turns on the channel 2 overload
LED. This line goes low when the A2 input assembly detects that the amplitude of the channel
2 input signal exceeded the set range. This line is only used in four channel analyzers.
OVLDCn Overload C — In a two channel analyzer, a low on this line turns on the channel 1 overload
LED. In a four channel analyzer, a low on this line turns on the channel 3 overload LED. This
line goes low when the A1 Input assembly detects that the amplitude of the channel 1 input
signal exceeded the set range or when the A2 Input assembly detects that the amplitude of the
channel 3 input signal exceeded the set range.
OVLDDn Overload D — In a two channel analyzer, a low on this line turns on the channel 2 overload
LED. In a four channel analyzer, a low on this line turns on the channel 4 overload LED. This
line goes low when the A1 Input assembly detects that the amplitude of the channel 2 input
signal exceeded the set range or when the A2 Input assembly detects that the amplitude of the
channel 4 input signal exceeded the set range.
ROW0 - ROW6 Row 0 - Row 7 — When the row lines are set high, the row line that remains low indicates that
a key in that row was pressed.
SRCn Source On — A low on this line turns on the source LED. This line goes low when the Source
is turned on.
Voltages and Signals Agilent 35670A
A13 Primary Keypad
9-22