To set up the external trigger test
Performance Test only
This test verifies that the Agilent 35670A meets its trigger specification for external
trigger level accuracy. In this test, a frequency synthesizer outputs a 1 kHz signal
to the external trigger input and a 12.8 kHz signal to channel 1. The analyzer
makes an accurate triggered measurement on channel 1 to verify the trigger level
and slope.
1 2ch Using an HP 3326A
1 4ch Using an HP 3326A
Agilent 35670A Verifying Specifications
To set up the external trigger test
3-35