Agilent Technologies 35670-90066 Stereo System User Manual


 
Program Controlled Test Equipment
This program automatically controls the instruments listed in the following
table using GPIB commands. If you use a test instrument other than those
shown in the table, the program prompts you to set the instrument state during
testing.
Test Equipment Program Controlled Model
AC Calibrator Fluke 5700A
Alternate
Fluke 5200A
Datron 4200, 4707, 4708
Frequency Synthesizer HP 3326A
Alternate
(2) HP 3325A/B
Digital Multimeter HP 3458A
Alternate
HP 3455A
HP 3456A
HP 3478A
Measurement Uncertainty
A table starting on page 3-56 lists the measurement uncertainty and ratio for
each performance test using the recommended test equipment. Except for the
External Trigger test, the ratios listed for the recommended test equipment
meet or exceed the measurement uncertainty ratio required by
U.S.
MIL-STD-45662A
. The table also provides a place to record the measurement
uncertainty and ratio for each performance test using equipment other than the
recommended test equipment. The table may be reproduced without written
permission of Agilent Technologies.
Operation Verification and Performance Tests
The operation verification tests give a high confidence level (>90%) that the
Agilent 35670A Dynamic Signal Analyzer is operating properly and within
specifications. The operation verification tests are a subset of the performance
tests. The operation verification tests should be used for incoming and
after-repair inspections. The performance tests provide the highest level of
confidence and are used to verify that the Agilent 35670A Dynamic Signal
Analyzer conforms to its published specifications. Some repairs require a
performance test to be done after the repair (see chapter 6, ‘’Replacing
Assemblies’’ in the
Agilent 35670A Service Guide for this information). The
following table lists the operation verification and performance tests.
Verifying Specifications Agilent 35670A
3-4