Agilent Technologies 35670-90066 Stereo System User Manual


 
Press the following keys and type in the printer address when the program
prompts you:
[
TEST CONFIG ]
[
PRINTER ADDRESS ]
[
PROCEDURE ]
[
OP_VERIFY ]or[PERFORMAN ]
[
STOP AFTER ]
[
LIMIT FAILURE ]or[NONE ]
[
RETURN ]
Press the following keys to start the test:
[
START TESTING ]
[
START BEGINNING ]
When you select [
START BEGINNING ], the data is written to a file on the disk and
printed only after all tests are done. When you select [
START MIDDLE ]or
[
ONE TEST ], the data is printed immediately after each measurement.
Follow the directions on the display.
Warning During the test, the program prompts you to change the test equipment
connections. Always turn the ac calibrator output to OFF or STANDBY before
changing test equipment connections. The ac calibrator can produce output
voltages that could result in injury to personnel.
The directions on the display briefly tell you how to connect test equipment. For
detailed illustrations of equipment setup, see the setup illustrations starting on page
3-13.
If you want to pause the program and return the Agilent 35670A Dynamic Signal
Analyzer to front panel control, press [
BASIC ]. To continue the program, press
[
BASIC ][DISPLAY SETUP ][LOWER ][RETURN ][CONTINUE ]. If you changed
any instrument setup states, press [
RESTART TEST ] instead of [ CONTINUE ]to
ensure accurate measurement results.
Agilent 35670A Verifying Specifications
To run the program in semiautomated mode
3-9