Agilent Technologies 35670-90066 Stereo System User Manual


 
To load the program
For information about the program’s softkeys, see the menu descriptions starting on
page 3-51.
Set the Agilent 35670A Dynamic Signal Analyzer’s power switch to off (
O ),
then connect the analyzer, test instruments, and printer using GPIB cables.
If you have the PC Style Keyboard, option 1CL, connect the keyboard to the
analyzer using the keyboard cable (see ‘’To connect the optional keyboard’’ in
chapter 2).
Insert the Semiautomated Performance Test Disk into the analyzer’s disk drive,
then set the power switch to on (
l ).
After the analyzer finishes its power-up calibration routine, press the following
keys:
[
Local/GPIB ]
[
SYSTEM CONTROLLR ]
[
System Utility ]
[
MEMORY USAGE ]
[
REMOVE WATERFALL ]
[
CONFIRM REMOVE ]
[
RETURN ]
[
MORE ]
[
SERVICE TESTS ]
[
PERFRMANC TEST ]
Now go to one of the following procedures to continue:
‘’To run the program in semiautomated mode’’
‘’To run the program without a printer’’
‘’To run the program in manual mode’’
Agilent 35670A Verifying Specifications
To load the program
3-7