To load the program
For information about the program’s softkeys, see the menu descriptions starting on
page 3-51.
•
Set the Agilent 35670A Dynamic Signal Analyzer’s power switch to off (
O ),
then connect the analyzer, test instruments, and printer using GPIB cables.
•
If you have the PC Style Keyboard, option 1CL, connect the keyboard to the
analyzer using the keyboard cable (see ‘’To connect the optional keyboard’’ in
chapter 2).
•
Insert the Semiautomated Performance Test Disk into the analyzer’s disk drive,
then set the power switch to on (
l ).
•
After the analyzer finishes its power-up calibration routine, press the following
keys:
[
Local/GPIB ]
[
SYSTEM CONTROLLR ]
[
System Utility ]
[
MEMORY USAGE ]
[
REMOVE WATERFALL ]
[
CONFIRM REMOVE ]
[
RETURN ]
[
MORE ]
[
SERVICE TESTS ]
[
PERFRMANC TEST ]
• Now go to one of the following procedures to continue:
•
‘’To run the program in semiautomated mode’’
•
‘’To run the program without a printer’’
•
‘’To run the program in manual mode’’
Agilent 35670A Verifying Specifications
To load the program
3-7