To run the program in semiautomated mode
You must have an GPIB printer connected to your system to run the program in
semiautomated mode. If you do not have a printer, see ‘’To run the program without a
printer’’ later in this chapter.
Press the following keys and when the program prompts you, type in the
information for the title page of the test record and press [
ENTER ]:
[
TITLE PAGE ]
[
TEST FACILITY ]
[
FACILITY ADDRESS ]
[
TESTED BY ]
[
REPORT NUMBER ]
[
CUSTOMER ]
[
MORE ]
[
TEMP ]
[
HUMIDITY ]
[
LINE FREQUENCY ]
[
RETURN ]
Press the following keys and when the program prompts you, type in the
equipment configuration information:
[
EQUIP CONFIG ]
[
AC CALIBRATO ]
[
SYNTH. 1 ]
[
SYNTH. 2 ] (If needed)
[
LOW-D OSCILLATO ] (If needed)
[
MULTIMETER ]
[
RETURN ]
The GPIB address is 100 x (interface select code) + (primary address). The interface
select code for the test equipment and printer is 7 (for example, if the primary address
is 8, the GPIB address is 708).
When entering the calibration due date, only four characters are displayed on the
screen. However, you can enter up to nine characters and they will be printed.
Verifying Specifications Agilent 35670A
To run the program in semiautomated mode