Agilent Technologies 35670-90066 Stereo System User Manual


 
Agilent 35670A Verifying Specifications
3-53
Test Configuration Menu Descriptions
Press [ TEST CONFIG ] to display the
test configuration and the following
softkeys:
[
Agilent 35670A ADDRESS ]
Prompts you to enter the GPIB address
for the Agilent 35670A Dynamic Signal
Analyzer.
The GPIB addresses equals 100
(interface select code) + (primary
address). The interface select code for
the printer and test equipment is 7 (for
example, if the primary address is 8, the
GPIB address is 708).
[
PRINTER ADDRESS ]
Prompts you to enter the GPIB address
for the printer. To disable the printer,
set the printer address to 0.
[
PROCEDURE ]
Prompts you to select the operation
verification procedure (OP_VERIFY) or
the performance test procedure
(PERFORMAN).
[
BEEPER ]
Toggles the beeper on and off. When
the beeper is on, the program beeps
approximately every 2 minutes while
waiting for you to follow the directions
on the display and press [
CONTINUE ].
[
RETURN ]
Returns to the ITM_35670A main
menu.
[
STOP AFTER ]
Prompts you to select stop after limit
failure, stop after each measurement, or
do not stop after a limit failure or
measurement. If [
Limit Failure ]is
selected, the program stops after the
failing measurement is displayed but
before it is printed. At this point you
can continue on and print the failing
measurement or restart the
measurement.