Agilent Technologies 35670-90066 Stereo System User Manual


 
Preparing the Analyzer for Use
This chapter contains instructions for inspecting and installing the
Agilent 35670A Dynamic Signal Analyzer. This chapter also includes
instructions for cleaning the screen, transporting and storing the analyzer.
DC Power Requirements
The analyzer can operate from a dc power source supplying a true range of 10.8
to 30.8 Vdc. With all options installed, power consumption is less than
200 VA. The following table shows typical current requirements at different
operating voltages for the standard two-channel analyzer and for the optional
four-channel analyzer.
Operating Typical Current
Voltage Standard 2 channel
Agilent 35670A
Optional 4 channel Agilent 35670A
12 Vdc 8.0 amps 11.0 amps
24 Vdc 4.0 amps 5.5 amps
AC Power Requirements
The analyzer can operate from a 47 to 440 Hz, single-phase, ac power source
supplying 90 to 264 Vrms. With all options installed, power consumption is
less than 350 VA.
Warning Only a qualified service person, aware of the hazards involved, should measure
the line voltage.
2-2