Agilent Technologies 35670-90066 Stereo System User Manual


 
Step 4. Check external trigger signal to the Analog assembly.
Set the power switch to off ( O ).
Remove the seven screws holding the rear panel to the analyzer and lean the
rear panel back until the A10 Rear Panel assembly is visible. Keep the
cables connected.
Set the power switch to on ( l ).
Change the frequency synthesizer’s amplitude to 2 Vp-p.
Set the oscilloscope as follows:
CH1 V/div
Input Impedance
CH1 Coupling
Time/div
Probe Atten
100 mV/div
1M
dc
200 µs/div
10
Connect the oscilloscope to A10 TP2 using a 10:1 probe.
Press the following keys:
[
Trigger ]
[
EXTERNAL TRIGGER ]
[
TRIGGER SETUP ]
[
EXT LEVEL USER ]
[
EXT RANGE +/ 10 ]
The oscilloscope should display a 370
40 mVp-p square wave.
If the signal is incorrect, the A10 Rear Panel assembly is probably faulty.
Set the oscilloscope to 300 mV/div.
Press [ EXT RANGE +/ 2 ].
The oscilloscope should display a 1.9
0.2 Vp-p square wave.
If the signal is incorrect, the A10 Rear Panel assembly is probably faulty.
If the external trigger signals are correct, the A5 Analog assembly is
probably faulty.
Troubleshooting the Analyzer Agilent 35670A
To troubleshoot trigger failures
4-66