Configuring Inputs
Configuring System Data Configuring APIConfiguring InputsConfiguring Security & Language
Figure 7-3 Effect of flow direction on milliamp outputs: 4 mA value = 0
Volume flow type Liquid • Select the desired volume flow type.
- If you select Liquid, only liquid volume units are available for measurement
- If you select Gas, only gas standard volume units are available for
measurement. See the section entitled Gas standard volume flow and
standard density.
• If you are using the enhanced density application or the petroleum
measurement application, select Liquid.
Volume units
(1)
l/s Select the desired unit of liquid volume flow. See Table 7-5. The liquid
volume totalizer and inventory will use the corresponding volume unit.
Volume low flow
cutoff
(1)
0.00000 l/s • Enter the liquid volume flow rate below which volume flow outputs and
displays will indicate zero flow. See the section entitled Cutoffs later in this
chapter.
• Milliamp outputs have an additional cutoff parameter. See Section 8.4.4.
Standard gas volume
units
(2)
SCFM • Select the desired unit of gas standard volume flow. See Table 7-6. The gas
standard totalizer and inventory will use the corresponding volume unit.
Standard gas volume
cutoff
(3)(2)
0.0000 SCFM Enter the volume flow rate below which volume flow outputs and displays will
indicate zero flow. See the section entitled Cutoffs later in this chapter.
• Milliamp outputs have an additional cutoff parameter. See Section 8.4.4.
Standard gas density
(2)
0.10000 g/cm
3
• Enter the standard density of the gas you are measuring. See the section
entitled Gas standard volume flow and standard density
(1) Displayed only if Volume Flow Type = Liquid.
(2) Displayed only if Volume Flow Type = Gas Standard.
(3) Available only on systems with the enhanced core processor.
Table 7-2 Flow parameters continued
Variable Default Description
Reverse
flow
(1)
20
12
4
x0
20
12
4
-x x0
Milliamp output configuration:
20 mA value = x
4 mA value = 0
To set the 4 mA and 20 mA values, see
Section 8.4.4.
Forward
flow
(2)
Zero flow
Reverse
flow
(1)
Forward
flow
(2)
Zero flow
Flow direction parameter:
Forward only
Flow direction parameter:
Reverse only
Negate/Forward only
20
12
4
-x x0
Reverse
flow
(1)
Forward
flow
(2)
Zero flow
Flow direction parameter:
Absolute value
Bidirectional
Negate/Bidirectional
(1) Process fluid flowing in opposite direction from flow direction arrow on sensor.
(2) Process fluid flowing in same direction as flow direction arrow on sensor.
-x
mA output
mA output