Diagnostics and Troubleshooting
22.7.4 Configuration alarms
Table 22-8 lists all alarms in the Configuration category, with descriptions, suggested user actions,
and related information.
Table 22-8 Using configuration alarms
Alarm
number
Maintenance
menu listing Description
Alarm severity
Affected
by fault
timeout User actionsDefault
User
config?
A006 Characterize
Meter
Master reset has
been performed.
Required
calibration values
are missing.
Fault Yes No Enter required values. See
Section 7.3.3.
A020 Cal. Factors
Missing
Master reset has
been performed.
Required
calibration values
are missing.
Fault Yes No Enter required values. See
Section 7.3.3.
A021 Sensor Type
Incorrect
K1 value missing
or incorrect, or
RTD data from
sensor is
incorrect.
Fault No No Verify characterization
parameters. See
Section 7.3.3.
A027 Security Breach (Custody transfer
only) Device has
been tampered
with.
Fault No No Weights and Measures
security seal has been
broken. Alarm can be
cleared by user, but
authorized procedure is
required to reestablish
security. Ensure that the
security switch is set to the
ON position. Contact Micro
Motion.
A032
(1)(2)
Meter Verify Fault Meter verification
in progress, with
outputs set to
fault.
Fault No No Allow the procedure to
complete. If desired, abort
the procedure and restart
with outputs set to last
measured value.
A032
(3)
Outputs Fixed Meter verification
in progress, with
outputs set to
Fault or Last
Measured Value.
Varies
(4)
Allow the procedure to
complete.
If desired, abort the
procedure and restart with
outputs set to Continue
Measurement.
A034
(3)
Meter Verify
Failed
The test ran to
completion but
the results were
not within limits.
Info Yes No Rerun the test. If the test fails
again, see Section 21.3.4.
A035
(3)
Meter Verify
Aborted
The test did not
run to
completion.
Info Yes No If desired, read the abort
code, see Section 21.3.4,
and perform the appropriate
action.