Configuration and Use Manual 275
Default Values and Ranges
Relay Installation Series 3000 MenusDefaultsSpecifications – Model 3100
Milliamp output
2
Fault indication: Condition Downscale
Fault indication: Setting 2.0000 mA 1.0 – 3.6 mA
Fault indication: Last
measured value timeout
0 sec 0.0 – 60.0 sec
Variable assignment Density
Calibration span: 20.0 mA 10.0000 g/cm
3
Calibration span: 4.0 mA 0.0000 g/cm
3
Damping seconds 0.0000
LRV Mass flow –200.000 g/s
Volume flow –0.200 l/s
Density 0.000 g/cm
3
Temperature –240.000 °C
Drive gain 0.000%
Gas standard volume flow –423.78 SCFM
External temperature –240.000 °C
External pressure 0.000 psi
URV Mass flow 200.000 g/s
Volume flow 0.200 l/s
Density 10.000 g/cm
3
Temperature 450.000 °C
Drive gain 100.000%
Gas standard volume flow 423.78 SCFM
External temperature 450.000 °C
External pressure 100.000 psi
Frequency
output
Flow source Mass flow
Scaling method Frequency = flow
Frequency 1,000.00 Hz 0.00091 –
10,000.00 Hz
Flow 16,666.66999 g/s
Max pulse width 277.2352 ms 0 – 277.2352 ms
Power Active
Polarity Active high
Fault indication Downscale
Last measured value timeout 0.0 sec 0.0 – 60.0 sec
RS-485
parameters –
HART protocol
Baud rate 9600
Parity Odd
Data bits 8
Stop bits 1
Polling address 0
Table E-1 Series 3000 default values and ranges continued
Type Setting Default Range Comments