Series 3000 MVD Transmitters and Controllers
Measurement Performance
Figure 21-9 Smart Meter Verification scheduler – Display
Figure 21-10 Smart Meter Verification scheduler – Communicator
21.4 Performing meter validation
To perform meter validation, measure a sample of the process fluid and compare the measurement
with the flowmeter’s reported value.
Use the following formula to calculate a meter factor:
Valid values for meter factors range from
0.8 to 1.2. If the calculated meter factor exceeds these
limits, contact Micro Motion customer service.
To enter the new meter factor, use the Inputs menu (see Figure 7-1).
Schedule Meter Verify
Verbleibende Zeit
Nächster Durchgang
Wiederholung Zeit
Auswahl
Maintenance Menu >
Meter Verification >
1 Next Run
2 Set Hrs Until Next Run
3 Set Recurring Hours
4 Turn Off Schedule
3
1 Run Meter Verification
2 View Test Results
3 Schedule Meter Verification
6 3
Online >
1 Overview >
3 Shortcuts >
6 Meter Verification
Online >
3 Service Tools >
4 Maintenance >
1 Routine Maintenance >
3 Meter Verification
NewMeterFactor ConfiguredMeterFactor
ExternalStandard
ActualTransmitterMeasurement
-----------------------------------------------------------------------------------
×=