Configuration and Use Manual 223
Diagnostics and Troubleshooting
Diagnostics and Troubleshooting Specifications – Model 3350/3700Specifications – Model 3300/3500Measurement Performance
A101 mA 1 Fixed Primary mA
output is
transmitting fixed
output level.
Info Yes
(5)
No Change Loop Current Mode
parameter. See
Section 22.11.7.
Exit from mA output trim.
See Section 16.5.
Exit from mA output
simulation. See
Section 16.4.4.
Check to see if the output
has been fixed via digital
communication.
A104 Cal in Progress Transmitter
currently
performing flow
or density
calibration.
Info Yes
(5)
No No action required.
A106 Burst Enabled Transmitter
configured for
burst mode.
Info Yes
(5)
No No action required.
A111 Freq. Output
Fixed
Frequency output
is transmitting
fixed output level.
Info Yes
(5)
No Exit from frequency output
simulation. See
Section 16.4.2.
A114 mA 2 Fixed Secondary mA
output is
transmitting fixed
output level.
Info Yes
(5)
No Exit from mA output trim.
See Section 16.5.
Exit from mA output
simulation. See
Section 16.4.4.
Check to see if the output
has been fixed via digital
communication.
A118 DO 1 Fixed Discrete output 1
transmitting fixed
output.
Info Yes
(5)
No Unfix discrete output 1. See
Section 16.4.1.
A119 DO 2 Fixed Discrete output 2
transmitting fixed
output.
Info Yes
(5)
No Unfix discrete output 2. See
Section 16.4.1.
A120 ED: Curve Fit Fail Configured
values for density
curves do not
meet accuracy
requirements.
Info No No Verify enhanced density
configuration. See the
enhanced density manual.
A122 DO 3 Fixed Discrete output 3
transmitting fixed
output.
Info Yes
(5)
No Unfix discrete output 3.
SeeSection 16.4.1.
A127 Batcher Needs
Reset
Tried to start a
batch without
resetting
previous batch.
Info Yes No See Section 22.7.3.
A128 Batcher Cannot
Start
Trans mitter
cannot start a
batch.
Info Yes No See Section 22.7.3.
Table 22-8 Using configuration alarms continued
Alarm
number
Maintenance
menu listing Description
Alarm severity
Affected
by fault
timeout User actionsDefault
User
config?