Emerson Series 3000 MVD Satellite Radio User Manual


 
Configuration and Use Manual 181
Measurement Performance
Diagnostics and Troubleshooting Specifications – Model 3350/3700Specifications – Model 3300/3500Measurement Performance
21.2.2 Meter validation and meter factors
Meter validation can be used with all Series 3000 systems (both transmitters and controllers). There
are no special requirements.
Meter validation compares a measurement value reported by the Series 3000 device with an external
measurement standard. Meter validation requires one data point.
Note: For meter validation to be useful, the external measurement standard must be more accurate
than the sensor. See the sensor’s product data sheet for its accuracy specification.
If the Series 3000 device’s mass flow, volume flow, or density measurement is significantly different
from the external measurement standard, you may want to adjust the corresponding meter factor. A
meter factor is the value by which the Series 3000 device multiplies the process variable value. The
default meter factors are
1.0, resulting in no difference between the data retrieved from the sensor and
the data reported externally.
Meter factors are typically used for proving the flowmeter against a weights and measures standard.
You may need to calculate and adjust meter factors periodically to comply with regulations.
21.2.3 Calibration
Calibration applies only to the Model 3500 and Model 3700 transmitters. There are no special
requirements.
Result data on display Pass/Fail/Abort for current test For all results stored on transmitter:
• Pass/Fail/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs
Result data with Communicator Pass/Caution/Abort for current test For all results stored on transmitter:
• Pass/Caution/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs
Comparison table for stored results
Comparison plot for stored results
Result data with ProLink II For all results stored on PC:
• Pass/Fail/Abort
Abort code (if relevant)
• Stiffness of the right and left pickoffs
• Test execution metadata
• Comparison graphs
• Test reports
• Data export and manipulation
capabilities
For all results stored on transmitter:
• Pass/Fail/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs
• Test execution metadata
• Comparison graphs
•Test reports
• Data export and manipulation
capabilities
Startup methods Manual Manual
Scheduler
Event
Discrete input
Table 21-2 Comparison of meter verification features and functions: original version vs. Smart Meter
Verification continued
Feature or function
Meter verification application
Original version Smart Meter Verification