Emerson Series 3000 MVD Satellite Radio User Manual


 
198 Micro Motion
®
Series 3000 MVD Transmitters and Controllers
Measurement Performance
Density calibration fluids
D1 and D2 density calibration require a D1 (low-density) fluid and a D2 (high-density) fluid. You
may use air and water. If you are calibrating a T-Series sensor, the D1 fluid must be air and the D2
fluid must be water.
For D3 density calibration, the D3 fluid must meet the following requirements:
Minimum density of 0.6 g/cm
3
Minimum difference of 0.1 g/cm
3
between the density of the D3 fluid and the density of water.
The density of the D3 fluid may be either greater or less than the density of water
For D4 density calibration, the D4 fluid must meet the following requirements:
Minimum density of 0.6 g/cm
3
Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the density of the
D3 fluid. The density of the D4 fluid must be greater than the density of the D3 fluid
Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the density of water.
The density of the D4 fluid may be either greater or less than the density of water
During the calibration, you must enter the density of the calibration fluid. For convenience, Tables
21-4 and 21-5 provide the density of air and water at different environmental conditions.
CAUTION
For T-Series sensors, the D1 calibration must be performed on air and the D2
calibration must be performed on water.
Table 21-4 Density of air
Pressure
in millibar
(inches of
mercury)
Temperature in °C and °F
10 °C
50 °F
15 °C
59 °F
20 °C
68 °F
25 °C
77 °F
30 °C
86 °F
35 °C
95 °F
40 °C
104 °F
45 °C
113 °F
50 °C
122 °F
850 (25.14) .0010 .0010 .0010 .0010 .0010 .0010 .0009 .0009 .0009
900 (26.62) .0011 .0011 .0011 .0010 .0010 .0010 .0010 .0010 .0009
950 (28.10) .0012 .0011 .0011 .0011 .0011 .0011 .0010 .0010 .0010
1000 (29.57) .0012 .0012 .0012 .0012 .0011 .0011 .0011 .0011 .0011
1050 (31.06) .0013 .0013 .0012 .0012 .0012 .0012 .0012 .0011 .0011