Series 3000 MVD Transmitters and Controllers
Operation Mode – Custody Transfer
Process variables
All flow rates Reported normally Reported as zero Reported normally Reported normally
Density and temperature Reported normally Reported normally Reported normally Reported normally
Process totalizers Increment normally
Cannot be stopped
Cannot be reset
unless flowrate is
zero. If any one is
reset:
• All others are
reset
automatically
• A test of the
display is
performed
automatically
Stopped
If any one is reset:
• All others are
reset
automatically
A test of the
display is
performed
automatically
Replaced by
transfer totalizers
Transfer totalizers
Increment normally
Replaced by
transfer totalizers
Transfer totalizers
stopped
Process inventories Increment normally
Cannot be stopped
Cannot be reset
Stopped
Cannot be reset
Increment normally
Cannot be stopped
Cannot be reset
Stopped
Cannot be reset
Transfer totalizers N/A N/A Rollover indicator:
• Cannot be cleared
manually
• Cleared
automatically
upon starting new
transfer
Rollover indicator
can be cleared
Transfer inventories N/A N/A Rollover indicator
cannot be cleared
Rollover indicator
can be cleared
Simulation mode N Y N Y
Alarm menu Normal Normal Alarm log password
required (if enabled)
Alarm log password
required (if enabled)
Table 19-1 Series 3000 behavior and available functions continued
Custody transfer (NTEP)
Custody transfer (OIML) and
custody transfer (OIML/batch)
Function Secured Unsecured Secured Unsecured