108 Micro Motion
®
Series 3000 MVD Transmitters and Controllers
Configuring Custody Transfer
Figure 14-1 Security menu for custody transfer (NTEP)
14.5 Configuring custody transfer (OIML) and custody transfer (OIML/batch)
To configure custody transfer (OIML) or custody transfer (OIML/batch):
1. Referring to Figure 14-2:
a. Set
World Area to OIML.
b. Set
Discrete Batch (if displayed) as desired:
Approved: The discrete batch application can be used for transaction measurement
and management.
Not Approved: The discrete batch application cannot be used for transaction
measurement and management.
c. Set
API or Enhanced Density (if displayed) as desired:
Approved: API or enhanced density process variables can be used as the transfer
variable.
Not Approved: API or enhanced density process variables cannot be used as the
transfer variable.
2. Ensure that the Series 3000 system, including all special applications (discrete batch,
petroleum measurement, enhanced density), is completely configured and tested. Verify the
following:
Volume Flow Type is set to Liquid. Gas standard volume flow cannot be used for OIML
custody transfer.
If you will use the discrete batch application for transaction measurement, verify that the
discrete batch flow source has been set to the process variable to be used for custody
transfer measurement (the transfer variable). Only the following process variables are
valid:
- Mass flow
- Volume flow (liquid)
- API corrected volume flow (if the petroleum measurement application was approved
in the previous step)
- ED net mass flow (if the enhanced density application was approved in the previous
step)
3. Perform a sensor zero. See Section 16.3.
World area
•NTEP
•OIML
Switch status
(1)
Security
Weights & Measures
(1) Read-only.