Emerson Series 3000 MVD Satellite Radio User Manual


 
Configuration and Use Manual 213
Diagnostics and Troubleshooting
Diagnostics and Troubleshooting Specifications – Model 3350/3700Specifications – Model 3300/3500Measurement Performance
A009 Xmtr Initializing Device is
performing
self-calibration
after power-up.
Fault Yes No No action required unless
alarm does not clear. If alarm
does not clear:
• Check wiring to sensor.
• Check power supply wiring.
• Make sure that the sensor
is completely full or
completely empty.
• Verify sensor configuration.
See Section 7.3.3.
• If A026 alarm is present,
check core processor
address. See Section 19.6.
A014 Transmitter Error Transmitter
failure for any of
several reasons.
Fault No No Cycle power to the
flowmeter.
Perform sensor coil and
resistance tests. See
Section 22.20.
The flowmeter might need
service. Contact Micro
Motion customer service.
A018 EEPROM (3000) Uncorrectable
checksum
mismatch.
Fault No No Cycle power to the
flowmeter.
The flowmeter might need
service. Contact Micro
Motion customer service.
A019 RAM Failure A RAM location
in the transmitter
cannot be written
to.
Fault No No Cycle power to the
flowmeter.
The flowmeter might need
service. Contact Micro
Motion customer service.
A022
(1)
CP Config Failure Uncorrectable
checksum
mismatch.
Fault No No Cycle power to the
flowmeter.
The flowmeter might need
service. Contact Micro
Motion customer service.
A023
(1)
CP Totals Failure Uncorrectable
checksum
mismatch.
Fault No No Cycle power to the
flowmeter.
The flowmeter might need
service. Contact Micro
Motion customer service.
A024
(1)
CP Program
Failure
Uncorrectable
checksum
mismatch.
Fault No No Cycle power to the
flowmeter.
The flowmeter might need
service. Contact Micro
Motion customer service.
A025
(1)
CP Boot Program
Fault
Uncorrectable
checksum
mismatch.
Fault No No Cycle power to the
flowmeter.
The flowmeter might need
service. Contact Micro
Motion customer service.
Table 22-4 Electronics alarms continued
Alarm
number
Maintenance
menu listing Description
Alarm severity
Affected
by fault
timeout User actionsDefault
User
config?