Agilent Technologies E4406A VSA Satellite Radio User Manual


 
Chapter 2 81
Programming Fundamentals
Using the Instrument Status Registers
Generating a Service Request To use the SRQ method, you must
understand how service requests are generated. Bit 6 of the status byte
register is the request service (RQS) bit. The *SRE command is used to
configure the RQS bit to report changes in instrument status. When
such a change occurs, the RQS bit is set. It is cleared when the status
byte register is queried using *SRE? (with a serial poll.) It can be
queried without erasing the contents with *STB?.
When a register set causes a summary bit in the status byte to change
from 0 to 1, the instrument can initiate the service request (SRQ)
process. However, the process is only initiated if both of the following
conditions are true:
The corresponding bit of the service request enable register is also
set to 1.
The instrument does not have a service request pending. (A service
request is considered to be pending between the time the
instruments SRQ process is initiated and the time the controller
reads the status byte register.)
The SRQ process sets the GPIB SRQ line true. It also sets the status
bytes request service (RQS) bit to 1. Both actions are necessary to
inform the controller that the instrument requires service. Setting the
SRQ line only informs the controller that some device on the bus
requires service. Setting the RQS bit allows the controller to determine
which instrument requires service.
If your program enables the controller to detect and respond to service
requests, it should instruct the controller to perform a serial poll when
the GPIB SRQ line is set true. Each device on the bus returns the
contents of its status byte register in response to this poll. The device
whose RQS bit is set to 1 is the device that requested service.
NOTE When you read the instruments status byte register with a serial poll,
the RQS bit is reset to 0. Other bits in the register are not affected.
NOTE If the status register is configured to SRQ on end-of-measurement and
the measurement is in continuous mode, then restarting a
measurement (INIT command) can cause the measuring bit to pulse
low. This causes an SRQ when you have not actually reached the
"end-of-measurement" condition. To avoid this:
1. Set INITiate:CONTinuous off.
2. Set/enable the status registers.
3. Restart the measurement (send INIT).