Chapter 5 313
Language Reference
SENSe Subsystem
Remarks: You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
Adjacent Channel Power—Spectrum Trace Control
[:SENSe]:ACP:SPECtrum:ENABle OFF|ON|0|1
[:SENSe]:ACP:SPECtrum:ENABle?
Turns on/off the measurement of the spectrum trace data when the
spectrum view is selected. (Select the view with DISPlay:ACP:VIEW.)
You may want to disable the spectrum trace data part of the
measurement so you can increase the speed of the rest of the
measurement data.
Factory Preset
and *RST: ON
Remarks: You must be in Basic, cdmaOne, iDEN mode to use this
command. Use INSTrument:SELect to set the mode.
History: Revision A.03.27 or later, in cdmaOne revision A.04.00
Adjacent Channel Power—Sweep Mode Resolution Bandwidth
[:SENSe]:ACP:SWEep:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:ACP:SWEep:BANDwidth|BWIDth[:RESolution]?
Sets the resolution bandwidth when using the spectrum analyzer type
sweep mode. See [:SENSe]:ACP:SWEep:TYPE.
Factory Preset
and *RST: Auto coupled.
Range: 1.0 kHz to 1.0 MHz
Resolution: 1.0 kHz
Step Size: 1.0 kHz
Default Unit: Hz
Remarks: You must be in the cdmaOne cdma2000, W-CDMA
(3GPP), or W-CDMA (Trial & Arib) mode to use this
command. Use INSTrument:SELect to set the mode.