Agilent Technologies E4406A VSA Satellite Radio User Manual


 
Chapter 5 359
Language Reference
SENSe Subsystem
M6 - subtracts 6 dB of fixed gain across the range
P0 to 24 - adds 0 to 24 dB of fixed gain across the range
Factory Preset
and *RST: AUTO
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Waveform - Query Aperture Setting
[:SENSe]:WAVeform:APERture?
Returns the waveform sample period (aperture) based on current
resolution bandwidth, filter type, and decimation factor. Sample rate is
the reciprocal of period.
Remarks: To use this command the appropriate mode should be
selected with INSTrument:SELect.
History: Version A.05.00 or later
WaveformNumber of Averages
[:SENSe]:WAVeform:AVERage:COUNt <integer>
[:SENSe]:WAVeform:AVERage:COUNt?
Set the number of sweeps that will be averaged. After the specified
number of sweeps (average counts), the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST: 10
Range: 1 to 10,000
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
WaveformAveraging State
[:SENSe]:WAVeform:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:WAVeform:AVERage[:STATe]?
Turn averaging on or off.
Factory Preset
and *RST: OFF
Remarks: To use this command, the appropriate mode should be