Agilent Technologies E4406A VSA Satellite Radio User Manual


 
Chapter 5 331
Language Reference
SENSe Subsystem
History: Version A.03.00 or later
Front Panel
Access:
FREQUENCY/Channel, CF Stepl
RF Power Commands
RF Port Input Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <rel_power>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the RF input attenuator. This value is set at its auto value if input
attenuation is set to auto.
Factory Preset
and *RST: 0 dB
12 dB for iDEN
Range: 0 to 40 dB
Default Unit: dB
Front Panel
Access:
Input, Input Atten
RF Port Input Attenuator Auto
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Select the RF input attenuator range to be set either automatically or
manually.
ON - Input attenuation is automatically set as determined by the
reference level setting.
OFF - Input attenuation is manually set.
Front Panel
Access:
Input/Output (or Input), Input Atten