314 Chapter 5
Language Reference
SENSe Subsystem
Adjacent Channel Power—Sweep Mode Resolution BW Control
[:SENSe]:ACP:SWEep:BANDwidth|BWIDth[:RESolution]:AUTO
OFF|ON|0|1
[:SENSe]:ACP:SWEep:BANDwidth|BWIDth[:RESolution]:AUTO?
Sets the resolution bandwidth to automatic, when using the spectrum
analyzer type sweep mode. See [:SENSe]:ACP:SWEep:TYPE.
Factory Preset
and *RST: ON
Remarks: You must be in the cdmaOne cdma2000, W-CDMA
(3GPP), or W-CDMA (Trial & Arib) mode to use this
command. Use INSTrument:SELect to set the mode.
Adjacent Channel Power—Sweep Mode Detection
[:SENSe]:ACP:SWEep:DETector[:FUNCtion] AAVerage|POSitive
[:SENSe]:ACP:SWEep:DETector[:FUNCtion]?
Selects the detector type when using the sweep mode. See
[:SENSe]:ACP:SWEep:TYPE.
Absolute average (AAVerage) - the absolute average power in each
frequency is measured across the spectrum
Positive - the positive peak power in each frequency is measured
across the spectrum
Factory Preset
and *RST: POSitive
Remarks: You must be in the cdma2000, W-CDMA (3GPP), or
W-CDMA (Trial & Arib) mode to use this command.
Use INSTrument:SELect to set the mode.
Adjacent Channel Power—Sweep Time
[:SENSe]:ACP:SWEep:TIME <seconds>
[:SENSe]:ACP:SWEep:TIME?
Selects a specific sweep time used to measure the reference (carrier)
channel. If you increase the sweep time, you increase the length of the
time data captured and the number of points measured. You might need
to specify a specific sweep speed to accommodate a specific condition in
your transmitter. For example, you may have a burst signal and need to
measure an exact portion of the burst.