Agilent Technologies E4406A VSA Satellite Radio User Manual


 
204 Chapter 5
Language Reference
CALCulate Subsystem
CHPower - no markers
EVMQpsk - markers available
PSTatistic - markers available
RHO - markers available
SPECtrum - markers available
WAVeform - markers available
Example:
Suppose you are using the Spectrum measurement. To position marker
2 at the maximum peak value of the trace that marker 2 is currently on,
the command is:
:CALCulate:SPECtrum:MARKer2:MAXimum
You must make sure that the measurement is completed before trying
to query the marker value. Use the MEASure or READ command before
using the marker command. This forces the measurement to complete
before allowing the next command to be executed.
Markers All Off on All Traces
:CALCulate:<measurement>:MARKer:AOFF
Turns off all markers on all the traces in the specified measurement.
Example: CALC:SPEC:MARK:AOFF
Remarks: The keyword for the current measurement must be
specified in the command. (Some examples include:
SPECtrum, WAVeform)
Front Panel
Access:
Marker, More, Marker All Off
Marker Function
:CALCulate:<measurement>:MARKer[1]|2|3|4:FUNCtion
BPOWer|NOISe|OFF
:CALCulate:<measurement>:MARKer[1]|2|3|4:FUNCtion?
Selects the type of marker for the specified marker. A particular
measurement may not have all the types of markers that are commonly
available.
The marker must have already been assigned to a trace. Use
:CALCulate:<measurement>:MARKer[1]|2|3|4:TRACe to assign a
marker to a particular trace.