224 Chapter 5
Language Reference
CALibration Subsystem
A valid service password needs to be entered prior to
sending this command.
Front Panel
Access:
System, Alignments, Align subsystem, Align 50 MHz
Reference
Enter Interactive Mode for Internal 50 MHz Amplitude
Reference Alignment
:CALibration:REF50:ENTer
Turns on the interactive mode for alignment of the internal 50 MHz
amplitude reference signal. Use CAL:REF50:ANOW to do the
alignment and CAL:REF50:EXIT to exit the interactive mode.
Remarks: You must be in the Service mode to use this command.
Use INSTrument:SELect.
A valid service password needs to be entered prior to
sending this command.
Front Panel
Access:
System, Alignments, Align subsystem, Align 50 MHz
Reference
Exit Interactive Mode for Internal 50 MHz Amplitude Reference
Alignment
:CALibration:REF50:EXIT
Turns off the interactive mode for alignment of the internal 50 MHz
amplitude reference signal. Use CAL:REF50:ENTer to turn the mode
on and CAL:REF50:ANOW to do the alignment immediately.
Remarks: You must be in the Service mode to use this command.
Use INSTrument:SELect.
A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
System, Alignments, Align subsystem, Align 50 MHz
Reference
Query the Absolute Level for the 50 MHz Amplitude Reference
:CALibration:REF50:LAST:ALEVel?
Query returns the last value of the absolute level of the 50 MHz
reference alignment.